ADAM-5018P 7-ch Thermocouple Input Module with Independent Input Range (ADAM-5018P)
ADAM-5018P is a 16-bit, 7-channel thermocouple input module that features programmable independent input ranges on each channel. It accepts millivolt inputs (±15 mV, ±50 mV, ±100 mV, ±500 mV), voltage inputs (±1V, ±2.5 V), current input (±20 mA, 4~20 mA, requires 120 ohms resistor) and thermocouple input (J, K, T, R, S, E, B). The module forwards the data to the host computer in engineering units (mV, V, mA or temperature °C). It also accepts high common mode up to 200 Vdc and has many protections built in, like over voltage, isolation voltage, filter and TVS/ESD function. An external CJC on the plug-in terminal is designed for accurate temperature measurement.
- Features
- Specs
- Resources
- Order Info
ADAM-5018P 7-ch Thermocouple Input Module with Independent Input Range Features
High Common Mode: 2000 VDC
Fault and Overvoltage Protection: Withstands over voltage up to ±35 V
Isolation Voltage: 3,000 VDC
Filter function: Yes
Built-in TVS/ESD Protection
Input Impedance: 20 MΩ. (Build-in 120 Ω Register for Current Input)
Channels: 7 differential with Independent Input Range
CMR @ 50/60 Hz: 92 dB min
Sampling Rate: 10 samples/sec. (total)
Accuracy Bandwidth: ±0.1% or better, 13.1 Hz @ 50 Hz, 15.72 Hz @ 60 Hz
ADAM-5018P 7-ch Thermocouple Input Module with Independent Input Range Specifications
| General |
| Certifications | CE |
| Connectors | 1 x Plug-in screw terminal (# 14 ~ 22 AWG) |
| Power Consumption | 0.63 W (max.) |
| Thermocouple Input |
| Accuracy Bandwidth | ±0.1% or better |
|   | 13.1 Hz @ 50 Hz |
|   | 15.72 Hz @ 60 Hz |
| Channels | 7 differential with Independent Input Range |
| CMR @ 50/60 Hz | 92 dB min |
| Input Impedance | 20 MΩ. (Build-in 120Ω Register for Current Input) |
| lnput Range and Types | >> |
| Thermocouple | J 0 ~ 760° C |
|   | K 0 ~ 1,370° C |
|   | T -100 ~ 400° C |
|   | E 0 ~ 1,000° C |
|   | R 500 ~ 1,750° C |
|   | S 500 ~ 1,750° C |
|   | B 500 ~ 1,800° C |
| Current mode | ±20 mA, 4 ~ 20 mA |
| Voltage | ±15 mV (0.2%), ±50 mV, |
|   | ±100 mV, ±500 mV |
|   | ±1 V, ±2.5 V |
| Resolution | 16-bit |
| Sampling Rate | 10 samples/sec. (total) |
| Span Drift | ±25 PPM/°C |
| Zero Drift | ±6 ìV/°C |
| High Common Mode | 200 VDC |
| Protection |
| Fault and Overvoltage Protection | Withstands overvoltage up to ±35 V |
| Isolation Voltage | 3,000 VDC |
| Filter function | Yes |
| Note |
| (1) | Not support ADAM-5510KW series |
External links for ADAM-5018P
Ordering information for ADAM-5018P 7-ch Thermocouple Input Module with Independent Input Range
| Product code | Product options |